Skip to content

Commit

Permalink
explain holding current on vertical XAFS table motors
Browse files Browse the repository at this point in the history
  • Loading branch information
bruceravel committed Jan 2, 2025
1 parent 0d9bccd commit 3375362
Show file tree
Hide file tree
Showing 8 changed files with 134 additions and 17 deletions.
Binary file added docs/_images/drive_current.png
Loading
Sorry, something went wrong. Reload?
Sorry, we cannot display this file.
Sorry, this file is invalid so it cannot be displayed.
Binary file added docs/_images/holding_current.png
Loading
Sorry, something went wrong. Reload?
Sorry, we cannot display this file.
Sorry, this file is invalid so it cannot be displayed.
Binary file added docs/_images/kafka_consumers.png
Loading
Sorry, something went wrong. Reload?
Sorry, we cannot display this file.
Sorry, this file is invalid so it cannot be displayed.
Binary file added docs/_images/timeout.png
Loading
Sorry, something went wrong. Reload?
Sorry, we cannot display this file.
Sorry, this file is invalid so it cannot be displayed.
135 changes: 125 additions & 10 deletions docs/details.rst
Original file line number Diff line number Diff line change
Expand Up @@ -226,14 +226,28 @@ pointed to in databroker. See:
+ `BMM/user_ns/detectors.py <https://github.com/NSLS-II-BMM/profile_collection/blob/master/startup/BMM/user_ns/detectors.py#L253>`__
+ `BMM/camera_device.py <https://github.com/NSLS-II-BMM/profile_collection/blob/master/startup/BMM/camera_device.py#L62-L164>`__

While this resembles a properly integrated camera and counting on the
``anacam`` object will get recorded in the database, the resource file
|nd| the jpeg image |nd| gets written to the local machine, thus is
not recoverable via Tiled. A future todo would be to make a caproto
IOC as a wrapper around ``fswebcam`` and give it proper authority to
write to storage.

.. admonition:: Update
:class: note

As of December 2024, this has been captured in BMM's ansible
configuration. Thus xf06bm-ws3 should always have this udev rule
available, even after a system upgrade or installation.



Pilatus 100K
------------

.. todo:: Need to flesh this out with explanatory text and screenshots

How files saving works
~~~~~~~~~~~~~~~~~~~~~~
How files saving works:

+ tiff files to /disk2
+ /disk2 is mounted on xf06bm-ioc1
Expand All @@ -242,14 +256,17 @@ How files saving works
pilatus, puilatus_tiff is helpful for testing tiff file writing,
which is used by IBM

Visualization:

+ Set two ROIs named "specular" (ROI1) and "yoneda" (ROI2)
+ XDI header to identify HDF5 file


Moving the detector between end stations
~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~
Moving the detector between end stations:

+ power cables (strip and detector)
+ ethernet cables
+ GN2 line
+ ethernet cable
+ GN2 line, note flow rate on meter in rack
+ grounding line

The NFS server might need to be restarted after rebooting. As root on
Expand All @@ -259,6 +276,7 @@ xf06bm-pilatus100k, do
/etc/init.d/nfsserver restart
.. todo:: Could this be in a smaller package. The rack is kinda huge.



Expand Down Expand Up @@ -396,16 +414,20 @@ table in the snapshot below.
-10 5E-6
====================== ==========

Note that the voltages are positive on the 7-element and negative on
the 4-element.

Temperature reading should be 1.5 V when the TEC is at proper
temperature.
Temperature reading on the 4-element should be 1.5 V when the TEC is
at proper temperature. Temperature should be 0.6 V on the 7-element
detector.

`Vortex SDD manual
<https://www.aps.anl.gov/files/download/DET/Detector-Pool/Spectroscopic-Detectors/Vortex_SDD/Vortex_ME4/Vtx-ME4%20Multi-El%20User%20Manual%20Rev.4.pdf>`__
(link to copy at APS detector pool).

There is a copy of the Vortex manual at BMM. Look in ``/nsls2/data3/bmm/legacy/products/ME7/``,
the file is called ``Vtx-Multi-El User Manual Rev 15.0_Oct 16, 2023.pdf``.
There is a copy of the Vortex manual at BMM. Look in
``/nsls2/data3/bmm/legacy/products/ME7/``, the file is called
``Vtx-Multi-El User Manual Rev 15.0_Oct 16, 2023.pdf``.


DM3 CAT6 Patch Panel
Expand Down Expand Up @@ -904,3 +926,96 @@ The homing procedure works in the sense of finding the negative limit,
then moving to a home position. But that home position seems to be
about 1/10 of the way between the negative limit and the
home-using-encoder-counts.

MC09 patch panel
----------------

Photo and explain...


.. _holding_current:

XAFS table holding current
--------------------------

The controller axes for the vertical motion of the XAFS table were
originally configured without a holding current. Over time, we
noticed that the XAFS table would lose its position by a couple
millimeters over the course of months. In an effort to combat this
drift, a holding current was applied to those three axes in
January 2025.

The three vertical table axes are channels 1, 2, and 3 on MC07, which
is a Geobrick Power PMAC.

This turned out to be a little complicated. It turns out that the
holding current is controlled by PLC7 in the Geobrick Power PMAC
controllers. That PLC uses several P-variables on the controller, as
explained in `the comment block at the top of the PLC code
<https://code.nsls2.bnl.gov/xf/06bm/iocs/xf06bm-ioc/mc07/-/blob/master/tpmacApp/pmc/xf06bma-mc07-plc07-power-down.pmc>`__.
(Note: that link requires logging onto the BNL VPN.)

Here is the text of that comment:

.. code-block:: text
; Note1: Geobrick coontrollers when killed effectively short the motor cables together providing an
; brake due to back EMF when the motor is rotated. Most axis can be safely killed without losing
; position, this is certainly the case for most lead screw drives (ball screws may require a
; holding current).
;
; Note2: When using this PLC make sure the standard kill PLC (usually PLC7, sometimes PLC3) is removed from the geobrick.
;
; Settings:
; P701-P708
; Define timeout period in milliseconds after which the axis will be powered
; down provided it has been idle for the whole period (note clock resolution below).
; Set to zero to leave the amp powered continously.
; Set to one for (almost) immediate power down on motor stop.
; Set to number of milliseconds for delayed power down after motor stops.
; Typically this should be set to a few seconds.
; P733-740
; Drive current & percentage for each axis.
; Set to zero for axis that are to be killed or if controller is not a geobrick (i.e. does not support Ixx77 amp current).
; For axis requiring a reduced holding current this contains ' normal_drive_current * 100 + power_down_percentage '
; Note the drive current is defined in milliamps.
; e.g. For motor axis 1 with 2000 milliamp drive current and 33% holding
; current define P733 as 200033
OK! So how do we make all that happen? Firing up PEWIN on the
Windows laptop connected to MC07, open the windows for viewing the
I-variables and the P-variables.

All the relevant parameters were set in PEWIN as shown in the figures
below. MC07 was then power cycled. After power cycling, those three
axes have a good strong holding current in place.

A holding current of 30% of drive current was selected because the
table is rather heavy. Jakub Wlodek's suggestion was "I usually go
with 10% unless it is lifting something heavy".


.. subfigure:: ABC
:layout-sm: ABC
:subcaptions: above
:gap: 8px
:name: fig-holding-current
:class-grid: outline

.. image:: _images/timeout.png

.. image:: _images/holding_current.png

.. image:: _images/drive_current.png

(Left) The P-variables window showing parameters P701-708. The
values P701, P702, and P703 are for axes 1, 2, and 3. The value of
5000 |nd| disable amplifier 5000 ms after a move |nd| are changed
to the value of 0, thus leaving the amplifiers enabled. (Middle)
The P-variables window showing parameters P733-740. The values
P733, P734, and P735 are for axes 1, 2, and 3. These are set to
supply a holding current of 30% of the drive current using the
specified formula. (Right) The I-variables window showing the
relevant parameter for axis 1. The I177 parameters is set to 1800
mA, as are I277 and I377. This value was used to compute the value
of the P733 - 735 parameters.
2 changes: 2 additions & 0 deletions docs/instruments.rst
Original file line number Diff line number Diff line change
Expand Up @@ -364,3 +364,5 @@ pitch, or roll.

+ Rotation stage mechanical drawing: :download:`JPG <_static/RT-2-11-DIMENSIONS.JPG>`



7 changes: 3 additions & 4 deletions docs/kafka.rst
Original file line number Diff line number Diff line change
Expand Up @@ -120,14 +120,13 @@ Once all three workers are running, you are ready to start |bsui| or


.. _fig-consumer:
.. figure:: _images/consumer_startup.png
:target: _images/consumer_startup.png
:width: 70%
.. figure:: _images/kafka_consumers.png
:target: _images/kafka_consumers.png
:width: 80%
:align: center

A terminal window in which the Kafka consumers have been started.

.. todo:: Update this figure



Expand Down
7 changes: 4 additions & 3 deletions docs/restore.rst
Original file line number Diff line number Diff line change
Expand Up @@ -78,7 +78,8 @@ To re-power the Xspress3 and its associated server:
switched on. It should be glowing red.
#. Press the front power button.
#. Once running, restart the relevant IOC on ``xf06bm-ioc2``. For the
seven element detector, use ``xs3-7-1``.
seven element detector, use ``xs3-7-1``. For the four element
detector, use ``xs3-4-1``.


Other IOCs
Expand Down Expand Up @@ -155,7 +156,7 @@ IOC name purpose
MC11 goniometer motors
MC12 goniometer motors
MC13 goniometer motors
mythen1k Mythen (in use??)
mythen1k Mythen (not in use. Nonfunctional, according to Oksana)
omega_i_series ??
onewire 1Wire temerature sensors near mono
piE625-M2 M2 piezo controller
Expand Down Expand Up @@ -269,7 +270,7 @@ XRD position is around 107,000.

.. note::

Never home M1, the collimating mirror. It is close enough to the
**Never** home M1, the collimating mirror. It is close enough to the
right position and should never be moved. In fact, there is no
reason to power up the motor controller.